TURBO
TEMP Temperature
Scale °Fahrenheit (F)
°Celsius (C)
Range
Predictive mode Max: 41.1° C; 106.0° F
Min: 33.6° C; 96.0° F
Monitor mode Max: 41.1° C; 106.0° F
Min: 26.7° C; 80.0° F
Monitor mode accuracy ± 0.1° C
± 0.2° F (when tested in a
calibrated liquid bath; meets
ASTM E1112, Table 1, in range
specified)
Predictive mode accuracy ± 1.0° F
± 0.6° C
Determination time approximately 10 seconds
Use only IVAC probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention
problems may occur unless IVAC probes and probe covers are used.
Refer to Appendix D for reorder codes.
IVAC
®
Patents
U.S. D300,728; D300,909
NELLCOR SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate 20 to 250 beats/min
Perfusion Range 0.03 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adults* 70 to 100% ±2 digits
Without Motion - Neonate* 70 to 100% ±3 digits
With Motion - Adults/Neo** 70 to 100% ±3 digits
Low Perfusion 70 to 100% ±2 digits